(b)(4).To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent additional information was requested, and the following was obtained: date of event: (b)(6) 2021.Event description: tpw32 is placed during a cardiac procedure, but the event occurred during post-op in icu.It was reported that one of the arterial pacing wires broke at the junction of the needle and wire.Patient was not being paced at the time event occurred.Date of event: (b)(6) 2021.When the event occurred: post-op in cicu.Patient additional information - patient age: (b)(6).No samples.The following information was requested, but unavailable: what is the lot number?.
|
(b)(4).Date sent to the fda: 09/13/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained: 1.What instruments were used on the blunt needle/temporary pacing wire during/after placement? how/where is the blunt needle being grasped?- everything was going ok in or during tpw placement and needle breakaway process.During the needle breakaway, a needle holder/grasper was used to hold a part which coming off the needle to release a pin.Then an exposed pin part was taped/secured seems to the patient¿s body in order to move patient from or to icu.2.How/where is the temporary pacing wire being grasped?- they used hands on the wire side during needle breakaway process.3.Do you know if we should still expect another sample with needle breakaway piece? if yes, please specify under what pc number.¿sample which they planned to ship to us was already sent under different file.At this time, we do not have any other samples on hold or in the process of shipping.- it was reported that there are no problems during insertion.They usually use a stay suture where the wire exits the skin, plus tegaderm is used on top to secure.No intentional tension is placed on the wire.The hospital is seeing the suture is breaking near the swage days after insertion while still in the icu.They use kits prepackaged from distributor which makes it difficult to track the lot number.Distributor uses re-sterilization on their surgical packs.
|