SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71441353
Device Problems Break (1069); Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2021
Event Type  malfunction  
Event Description
It was reported that after a ukr, a jrny uni tibial impactor, had a reprocessing problem and it was not reprocessed by the sterile services, after it was used in a surgery performed in (b)(6), the device was broken and cracked.Surgery was cancelled as consequence of this problem.Patient was not injured.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned journey uni tibial impactor confirms the plastic bumper has a piece missing.This piece was not returned with the device.Also the bumper is cracked.The device shows signs of significant use and wear.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D8 and d9: device returned.G2: report source update.G4: add 510k code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY UNI TIBIAL IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12250814
MDR Text Key264271987
Report Number1020279-2021-06132
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010588050
UDI-Public03596010588050
Combination Product (y/n)N
PMA/PMN Number
K113038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441353
Device Catalogue Number71441353
Device Lot Number18HM01767
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Date Manufacturer Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-