SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71441353 |
Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/16/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that after a ukr, a jrny uni tibial impactor, had a reprocessing problem and it was not reprocessed by the sterile services, after it was used in a surgery performed in (b)(6), the device was broken and cracked.Surgery was cancelled as consequence of this problem.Patient was not injured.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection of the returned journey uni tibial impactor confirms the plastic bumper has a piece missing.This piece was not returned with the device.Also the bumper is cracked.The device shows signs of significant use and wear.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D8 and d9: device returned.G2: report source update.G4: add 510k code.
|
|
Search Alerts/Recalls
|
|
|