• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH- AT ARTIS PHENO; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH- AT ARTIS PHENO; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10849000
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/23/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis pheno system.During an interventional procedure, the user reported a sudden dark screen on the large display.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The detailed investigation showed that the problem was due to a hardware defect.The optical transmitter for transferring the image data was defective.This was revealed by the examination at the factory.Such a defect can only be rectified by service intervention and replacement of the affected component.The affected component was replaced during service intervention and the system functions again as specified.The occurrence rate of the aforementioned error pattern and the spare parts consumption of the affected part were checked.A possible error accumulation or even a systematic error, which would lead to a corrective action of the installed base, could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS PHENO
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key12258358
MDR Text Key264479490
Report Number3004977335-2021-90476
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869046877
UDI-Public04056869046877
Combination Product (y/n)N
PMA/PMN Number
K201156
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10849000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/24/2021
Initial Date FDA Received08/02/2021
Supplement Dates Manufacturer Received10/14/2021
Supplement Dates FDA Received10/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-