• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MCGRATH MAC; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MCGRATH MAC; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 301-000-000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problems Low Oxygen Saturation (2477); Unspecified Tissue Injury (4559)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Title: comparative efficacy of c-mac® miller videolaryngoscope versus mcgrath® mac size ¿1¿ videolaryngoscope in neonates and infants undergoing surgical procedures under general anesthesia: a prospective randomized controlled trial source: cmac miller vs.Mcgrath for intubation in infants.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the literature, the device was difficult to use secondary to fogging during intubation.It was also reported that the device had poor camera resolution and lack in the ease of intubation due to the plastic blade which made it difficult to apply force over the intraoral tissue.Another device was used for successful intubation.Few minor complication was noted like mucosal bleeding, desaturation and minor mucosal injury.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.It was reported that the display was defective.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH MAC
Type of Device
LARYNGOSCOPE, RIGID
MDR Report Key12264389
MDR Text Key266602354
Report Number9612501-2021-01272
Device Sequence Number1
Product Code CCW
UDI-Device Identifier10884521776494
UDI-Public10884521776494
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 10/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number301-000-000
Device Catalogue Number301-000-000
Was Device Available for Evaluation? No
Date Manufacturer Received09/30/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-