• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI-LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI-LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 273715
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/09/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Event Description
It was reported by the user facility that product is leaking.Illisisble writing fuite de solution.(product leaking).
 
Manufacturer Narrative
The failure mode of leakage (pre-activation) was verified based on photograph received for analysis showing the orange tinted foam tip and lidding.Ampoules are made from onion tubing skin borosilicate type i glass, which are design to break when pressure is applied to activate applicator at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator, breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass, it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handing.Production record review was completed for batch/lot 0322656 and no non-conformance was noted during the manufacturing of this lot.No further actions are required at this time.This failure will continue to be tracked and trended.H3 other text: see narrative below.
 
Event Description
It was reported by the user facility that product is leaking.Illisisble writing fuite de solution.(product leaking).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP HI-LITE ORANGE
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION 213, LLC 0113
1550 northwestern dr
el paso TX 79912
MDR Report Key12278693
MDR Text Key265159253
Report Number3004932373-2021-00357
Device Sequence Number1
Product Code KXG
Combination Product (y/n)Y
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number273715
Device Lot Number0322656
Initial Date Manufacturer Received 07/09/2021
Initial Date FDA Received08/04/2021
Supplement Dates Manufacturer Received09/02/2021
Supplement Dates FDA Received09/02/2021
Patient Sequence Number1
Patient Outcome(s) Other;
-
-