• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRICURE, INC. EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRICURE, INC. EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX Back to Search Results
Model Number CDK-1413
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Stroke/CVA (1770)
Event Date 07/03/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4) the epi-sense device was not returned for evaluation and a device history review was unable to be completed as the relevant lot number for the epi-sense device was not reported or able to be subsequently ascertained.
 
Event Description
It was reported that a patient underwent a staged hybrid ablation procedure, with the surgical epicardial portion performed on (b)(6) 2021 and the endocardial rf catheter portion on (b)(6) 2021.Both procedures were uneventful and the devices worked as expected.On (b)(6) 2021, the patient suffered a stroke.Catheter thrombectomy was performed without issue and patient is currently recovering.Historical data and the nature of the event dictate that this event is more likely associated with the endocardial catheter ablation.However, atricure cannot rule out potential contribution from the epicardial ablation.There was no reported device malfunction, and this event was the result of a procedural complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX
Type of Device
EPI-SENSE GUIDED COAGULATION SYSTEM WITH VISITRAX
Manufacturer (Section D)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer (Section G)
ATRICURE, INC.
7555 innovation way
mason OH 45040
Manufacturer Contact
john ehlert
7555 innovation way
mason, OH 45040
5136448220
MDR Report Key12284323
MDR Text Key265341756
Report Number3011706110-2021-00037
Device Sequence Number1
Product Code OCL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193162
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCDK-1413
Device Catalogue NumberCDK-1413
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/06/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
-
-