• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Insufficient Information (4580)
Event Date 07/09/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported disengagement failure of a perforator during surgery causing injury.Hemostasis was performed and the procedure was completed with no surgical delay.It is unknown the drill manufacturer, or if the perforator clicked in place on the drill.It is also unknown if the recommended spring tests were performed between each burr hole.
 
Event Description
N/a.
 
Manufacturer Narrative
Unique device identification (udi): (b)(4) or (b)(4).Per the reporter, the codman disposable perforator was not being returned for evaluation, and no lot number information has been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
MDR Report Key12289048
MDR Text Key265428557
Report Number3014334038-2021-00156
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/16/2021
Initial Date FDA Received08/06/2021
Supplement Dates Manufacturer Received08/31/2021
Supplement Dates FDA Received09/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-