Device Problems Use of Device Problem (1670); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  Injury  
Manufacturer Narrative
The reported event of "left ventricular outflow tract (lvot) turbulence" and "thought it might be a result of an improperly sized implant" could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that on (b)(6) 2021, a 29 mm epic stented porcine heart valve w/flexfit system was successfully implanted.After the implant procedure, during transesophageal echocardiography (tee), the device was observed to be causing left ventricular outflow tract (lvot) turbulence.The surgeon was not comfortable with this and thought it might be a result of an improperly sized implant.The patient was placed back onto bypass and the valve was removed and replaced with a 31 mm epic stented porcine heart valve w/flexfit system to resolve the event.Tee was performed, and it was determined that this was a correct size.There was no clinically significant delay in the procedure and the patient remained hemodynamically stable throughout.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC STENTED TISSUE VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
Manufacturer (Section G)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR   31310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12291249
MDR Text Key265571392
Report Number3001883144-2021-00114
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/09/2021
Initial Date FDA Received08/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age30 YR
-
-