SMITH & NEPHEW, INC. FAST-FIX 360 CURVED NDL DELIVERY SYS; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72202468 |
Device Problem
Premature Activation (1484)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: case (b)(4).
|
|
Event Description
|
It was reported that, during a meniscus repair, after deploying t1 t2 deployed prematurely.All ts were removed.The procedure was completed with a s+n back up device.No significant delay and no patient injury or other complications were reported.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Manufacturer Narrative
|
H3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product.No suture and anchors were returned.The needle overmold assembly was bent.The depth limiter button was in the default position.The actuator tip was near the top of the deployment channel.The actuator tip was twisted 90 degrees within the deployment channel.A functional evaluation was conducted.The actuator tip would not click into either position.The complaint was confirmed.Factors, which could have contributed to the complaint event, include an application of unintended inappropriate or excessive force to the device.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.No containment or corrective actions are recommended at this time.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection was performed on the product.No suture and anchors were returned.The needle overmold assembly was bent.The depth limiter button was in the default position.The actuator tip was near the top of the deployment channel.The actuator tip was twisted 90 degrees within the deployment channel.A functional evaluation was conducted.The actuator tip would not click into either position.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed and the root cause was associated with unintended use of the device.Factor that may have contributed to the reported event include inadvertently bending of the needle/overmold assembly.No containment or corrective actions are recommended at this tim internal complaint reference: (b)(4).B5: event description updated.
|
|
Event Description
|
It was reported that, during a meniscus repair, after deploying the t1 implant., the t2 implant deployed prematurely.All ts were removed.The procedure was completed with a s+n back up device.No further complications were reported.
|
|
Search Alerts/Recalls
|
|
|