Model Number 222239 |
Device Problem
Contamination (1120)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that while using 2 bd bbl¿ chromagar¿ orientation and bbl¿ trypticase¿ soy agar (tsa ii)-i plate¿ contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: it was reported that reported contamination.
|
|
Manufacturer Narrative
|
H6: investigation summary: during manufacturing of material 222239, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 1152672 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include physical attribute and bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All physical attribute and bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is.Tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and other complaints have been taken on batch 1152672.Retention samples from batch 1152672 were not available for inspection.Five photos were received for investigation.Three photos each show the bottom of a taped plate from batch 1152672 (time stamps 0959 and 1832) with what appears to be microbial growth in the tsa with 5% sheep blood medium.The other two photos each show two taped plates on a light bed.No returned samples were received for investigation of this complaint.This complaint has been confirmed.A trend was identified for contamination and investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) 3076308 has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.
|
|
Event Description
|
It was reported that while using 2 bd bbl¿ chromagar¿ orientation and bbl¿ trypticase¿ soy agar (tsa ii)-i plate¿ contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: " it was reported that reported contamination.".
|
|