• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Use of Device Problem (1670); Capturing Problem (2891); Device Dislodged or Dislocated (2923)
Patient Problems Discomfort (2330); Heart Failure/Congestive Heart Failure (4446)
Event Date 07/19/2021
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented for routine follow up complaining of malaise.In-clinic device interrogation revealed an increase in left ventricular capture threshold due to minor pull back of the left ventricular lead.Programming interventions were initially made, however the patient continued to experience heart failure symptoms.The physician decided to perform a revision but was unable to reposition the lead due to lack of guidewire support.The lead was explanted and replaced.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12295202
MDR Text Key265628217
Report Number2017865-2021-27781
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000107625
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 DA
Patient Weight80
-
-