This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d4, g3, g6, h2, h3, h4, h6, h10 the event was confirmed with product received.A g7 finned 3 hole shell 52e, part # 110017103 from lot 7001244, was returned and evaluated against the complaint.Visual inspection found the hex feature of the remaining screw plug to be stripped.The apical plug was also returned but exhibits no signs of damage or stripping.A small scratch was observed near the center screw plug hole but the shell is in good overall condition.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|