Model Number DM0210FAA
Device Problem Device Remains Activated (1525)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.This dm0210faa easydrill cranial perforator with lot number 1363/20 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a craniotomy procedure for a subdural hematoma of a trauma patient, the perforator did not stop.It was reported that there was no additional intervention performed and the procedure was completed using a competitive perforator.On follow-up, it was reported that there was a short delay (less than 3 minutes) due to retrieval and preparation of a codman perforator to complete the procedure.It was also confirmed that there was no patient or staff impact and there was no further information that can be provided regarding the status of the patient post-surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR  09980-39
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR   09980-39
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--39
BR   09980-39
140575722
MDR Report Key12304344
MDR Text Key266006701
Report Number1625507-2021-00271
Device Sequence Number1
Product Code HBF
UDI-Device Identifier07898959543180
UDI-Public07898959543180
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDM0210FAA
Device Catalogue NumberDM0210FAA
Device Lot Number1363/20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/14/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight88
-
-