Model Number AIA-2000 ST |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed noise by listening to the instrument and reproduced error by running a patient sample.Fse noticed friction from tray when rotating at high-speed causing squeaking sound.Fse removed rotor table, cleaned, and lubricated.Fse verified operation in mainte program, no further loud sounds.Fse ran daily check, qc, and patients¿ samples without error and within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The most probable cause of the reported event was due to friction from the tray causing incubator noise.
|
|
Event Description
|
The customer called to report a noise that might be coming from the incubator on the aia-2000 instrument.The tss asked her to open up the hood and perform an all set home operation.The customer stated the noise occurs without any error message when the test cup is being moved into the incubator.The instrument is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for luteinizing hormone (lhii), follicle stimulating hormone (fsh) and intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|