MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed noise by listening to the instrument and reproduced error by running a patient sample.Fse noticed friction from tray when rotating at high-speed causing squeaking sound.Fse removed rotor table, cleaned, and lubricated.Fse verified operation in mainte program, no further loud sounds.Fse ran daily check, qc, and patients¿ samples without error and within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The most probable cause of the reported event was due to friction from the tray causing incubator noise.
 
Event Description
The customer called to report a noise that might be coming from the incubator on the aia-2000 instrument.The tss asked her to open up the hood and perform an all set home operation.The customer stated the noise occurs without any error message when the test cup is being moved into the incubator.The instrument is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for luteinizing hormone (lhii), follicle stimulating hormone (fsh) and intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12308533
MDR Text Key269272866
Report Number3004529019-2021-00002
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-