|
Model Number AIA-900 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineering (fse) followed up with the customer over the phone to resolve reported event.Customer confirmed and reproduced reported error.While troubleshooting with the customer, the seal breaker assembly/finger screw was found loose causing seal on cup not to break causing error 2160.The customer was instructed to tightened seal breaker assembly/finger screw to correct and prevent further errors.The customer confirmed the aia-900 analyzer is operating as expected by running samples without error.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-900 operator's manual under section 12: flags and error messages states the following: [2160] c.Transfer not detected cup cause: the cup sensor s063 failed to detect the cup before it was grasped.Action: please contact the tosoh local representatives.Check s063, the cup pickup position, and the cup pickup operation.The most probable cause of the reported event is due to loose seal breaker finger screw.
|
|
Event Description
|
A customer reported getting error message "2160 c transfer not detected cup" on the aia-900 analyzer.The customer rebooted analyzer but error persisted.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), follicle stimulating hormone (fsh), luteinizing hormone (lhii) and estradiol (e2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|