Model Number AB-5100R |
Device Problems
No Audible Prompt/Feedback (2282); Impedance Problem (2950)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 07/14/2021 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced sound quality issues.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed external visual inspection.The device passed the photographic imaging inspection.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.Failure analysis was able to confirm the device failed due to fluctuating impedances, however the root cause of the problem could not be determined.No corrective action can be implemented.Advanced bionics will continue to monitor for failure mode of this type and will implement corrective action if necessary.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|