Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Brain Injury (2219)
Event Date 07/07/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the bit/tip of the perforator did not stop after perforation of the cranial bone.It did not disengage and tore the dura matter.Medical staff had to suture the dura matter and it was reported no consequences for the patient.
 
Manufacturer Narrative
The perforator was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
MDR Report Key12311804
MDR Text Key266152419
Report Number3014334038-2021-00162
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ64N95
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/22/2021
Initial Date FDA Received08/12/2021
Supplement Dates Manufacturer Received09/01/2021
Supplement Dates FDA Received09/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age22 YR
-
-