Catalog Number EX060403C
Device Problem Fracture (1260)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/19/2021
Event Type  Injury  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 02/2023).
 
Event Description
It was reported that during a stent procedure in the superficial femoral artery, the distal end of the stent allegedly fractured during insertion.Reportedly, the device was not removed.The current patient status is unknown.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation and image was not provided.Therefore, the investigation is inconclusive for the reported issue.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.Holding and handling of the system throughout deployment was found sufficiently described; in particular the instruction for use state: 'deployment of the stent is complete when the proximal stent radiopaque markers appose the vessel wall and the sheath radiopaque zone is proximal to the proximal stent radiopaque markers.' the instruction for use further state: 'predilitation of the lesion should be performed using standard techniques.' h10: d4 (expiry date: 02/2023), g3.H11: h6 (method).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a stent procedure in the superficial femoral artery the distal end of the stent allegedly fractured during insertion.Reportedly, the device was not removed.The current patient status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12317727
MDR Text Key266369096
Report Number9681442-2021-00440
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEX060403C
Device Lot NumberANFP1008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient SexMale
-
-