DORNOCH DUO FLUID CART WITH POWER IV, REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Decrease in Suction (1146); Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined that the unit had low suction and the inline filters and hepa filter were clogged due to smoke intake.The inline filters and hepa filter were replaced and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item.However, as the device is only assigned a unique serial number, an additional review could not be performed.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
It was reported the unit has weak suction after surgery.Upon tech inspection, both in-line filters and hepa filter were dirty from the account pulling smoke into the cart.No adverse events were reported as a result of this malfunction.
|
|
Search Alerts/Recalls
|
|
|