Model Number DTBA1D1
Device Problems Signal Artifact/Noise (1036); Electromagnetic Interference (1194); Over-Sensing (1438)
Patient Problem Chest Pain (1776)
Event Date 06/09/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 1258t lead implanted: (b)(6) 2014.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented to the emergency department experiencing chest pain.It was noted that the right atrial (ra) lead and right ventricular (rv) lead appears to be oversensing on atrial tachycardia/atrial fibrillation (at/af) and high-ratenon-sustained episodes.It was also noted that the cardiac resynchronization therapy defibrillator (crt-d) could be experiencing electromagnetic interference and there is a possibility that the rv lead had t-wave oversensing (twos).The device and leads remain in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12318709
MDR Text Key266409235
Report Number3004209178-2021-12374
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169007246
UDI-Public00643169007246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/28/2015
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2021
Date Device Manufactured04/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, 694965 LEAD
Patient Age68 YR
Patient Weight70
-
-