• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MERLIN PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MERLIN PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 3650
Device Problems Program or Algorithm Execution Problem (4033); Program or Algorithm Execution Failure (4036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer report number: 2017865-2021-28138.During capture threshold testing, the physician released the button to terminate the testing as expected, however, the device continued to decrement, resulting in a loss of capture.The test continued until completion.The patient had an intrinsic rhythm, so the loss of capture did not result in any patient consequences or adverse events.
 
Manufacturer Narrative
The device was subject to the decrement test field action issued by abbott on 10 march 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12319565
MDR Text Key266443668
Report Number2017865-2021-28136
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-03/10/22¿001¿C
Patient Sequence Number1
Treatment
GALLANT HF; GALLANT HF
-
-