As per a manufacturer incident report we received from the factory in (b)(6): during hysteroscopy, the patient had an air embolism, caused in a sudden desaturation and a heart rhythm disorder.As a medical intervention, the patient had to be intubated.
|
The device was sent back to the manufacturer.The investigation was performed on 2021-08-18.According to the results of the log files, no malfunction or non-conformity could be found.In addition, the device passed all functional tests as per factory specifications.The root cause of the air embolism is most likely that the tubing set was not primed prior to its usage.As written in the instructions for use (document id 96116041d, version 4.0, chapter 6.12), it is written that a functional check and the de-airing process must be performed prior to each application.
|