MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2021
Event Type  malfunction  
Manufacturer Narrative
The customer reported a complaint of the autopulse platform (serial # (b)(4)) was missing pixels on the liquid crystal display (lcd) and was not readable was confirmed during the functional testing.The root cause for the reported complaint was due to a defective lcd, likely attributed to normal wear and tear.The autopulse platform was manufactured in march 2016 and has reached its expected service life of 5 years.During visual inspection, unrelated to the reported complaint, noticed the front and bottom enclosures have multiple cracks in the screw well area.The observed physical damages are appeared to be the characteristics of harsh impact, likely caused by user mishandling such as a drop.The autopulse passed the initial functional test without any error.However, the lcd was observed with missing pixels, thus confirming the reported complaint.The lcd was replaced to remedy the display issue.Following service, the autopulse platform passed the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there was no similar complaint reported for autopulse platform with sn (b)(4).
 
Event Description
During shift check, the autopulse platform (serial # (b)(4)) was missing pixels on the liquid crystal display (lcd) and was not readable.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key12322303
MDR Text Key266625661
Report Number3010617000-2021-00776
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-