OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to okr.Okr checked the subject device and found that the reported phenomenon was duplicated due to the failure of front panel unit.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the service department of olympus (b)(4) (okr), it was found that panel display failure which was poor led number display occurred.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The cause of the indicated phenomenon was a failure of the led element of the front panel unit.We couldn't identify the cause of the led element failure.Since 4 years and 11 months have passed since manufacture, it is presumed that an accidental device malfunction.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12322313
MDR Text Key266764165
Report Number8010047-2021-10239
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/15/2021
Supplement Dates Manufacturer Received08/26/2021
Supplement Dates FDA Received09/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-