• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8537
Device Problems Fluid/Blood Leak (1250); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a clearlink system continu-flo solution set was breaking off in a non baxter extension set.It was further reported that when disconnecting, the male luer breaks off into the female luer of the non baxter extension set.This resulted in an isolyte leak and blood backing up in the intravenous (iv) line.This issue was identified during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.  visual inspection was performed to the photograph using the naked eye which revealed that the male luer rigid component gets stuck into female luer connector of b braun filtered extension tubing.The reported condition was verified.  by the nature of the sample, no additional tests were performed.  the cause of the condition could not be determined.  a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12324423
MDR Text Key266650318
Report Number1416980-2021-05019
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048970
UDI-Public(01)00085412048970
Combination Product (y/n)Y
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8537
Device Lot NumberR21D05048
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/21/2021
Initial Date FDA Received08/16/2021
Supplement Dates Manufacturer Received09/15/2021
Supplement Dates FDA Received09/17/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BRAUN EXTENSION SET; ISOLYTE; BRAUN EXTENSION SET; ISOLYTE
-
-