Catalog Number 2C8537 |
Device Problems
Fluid/Blood Leak (1250); Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the tubing of a clearlink system continu-flo solution set was breaking off in a non baxter extension set.It was further reported that when disconnecting, the male luer breaks off into the female luer of the non baxter extension set.This resulted in an isolyte leak and blood backing up in the intravenous (iv) line.This issue was identified during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation. visual inspection was performed to the photograph using the naked eye which revealed that the male luer rigid component gets stuck into female luer connector of b braun filtered extension tubing.The reported condition was verified. by the nature of the sample, no additional tests were performed. the cause of the condition could not be determined. a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|