The functional and visual check of the instrument prior to the repair showed that it was running out of specification.During the test run it was noticeable that bur was wobbling, and the bearings have been vibrating.This indicated already that the ball bearings have been worn out.During disassembling of the instrument, they fell apart due to wear and tear.To avoid such issues the instruction for use contains already following warnings, notesand advises: 2.2 technical condition: a damaged device or components could injure patients, users and third parties.Only operate devices or components if they are undamaged on the outside.Check that the device is working properly and is in satisfactory condition before each use.Have parts with sites of breakage or surface changes checked by the service.Safety checks may only be performed by trained service personnel.If the following defects occur, stop working and have the service personnel carry out repair work: malfunctions, damage, irregular running noise, excessive vibration, overheating, dental bur or diamond grinder is not firmly locked in the handpiece.2.5 service and repair: following expiry of the warranty, check the cutter/grinder holding system and quick stop once a year.Kavo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.Define the service interval depending on the frequency of use.
|