Model Number AIA-2000 ST
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error log, but could not reproduce error.The customer cleared stuck cups before fse was onsite.Fse removed waste chute and cleaned dried serum buildup.Fse successfully performed mainte program, daily check and quality control run without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2163] y-axis cup transfer cup release failure.Cause: the cup-gripping sensor detected a cup after the cup release operation.Solution: contact tosoh service center or local representatives.The most probable cause of the reported event is due to cups stuck and serum buildup in waste chute.
 
Event Description
A customer reported getting error "2163 y-axis cup transfer cup release failure" on the aia 2000 instrument.Prior to calling customer saw jammed cups in the waste chute and tried to clean them out.The error re-occurred.Technical support specialist (tss) asked if the customer could clean waste chute but customer would like on site service.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth), follicle stimulating hormone (fsh) and luteinizing hormone (lhii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12335000
MDR Text Key267044795
Report Number3004529019-2021-00008
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/02/2021
Initial Date FDA Received08/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse