Model Number 383400
Device Problems Inaccurate Flow Rate (1249); Excess Flow or Over-Infusion (1311)
Patient Problem Feeding Problem (1850)
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event Description
The customer reported that the device was overfeeding.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed.The unit was tested based on an accuracy test using 125ml of water.The delivered rate was as expected.The unit was within tolerance specifications.The unit¿s software was reinstalled as a preventative measure.The battery was replaced due to being out of date.The gasket and tie were replaced due to opening the unit.The pip pcb was replaced due to corrosion.The serial number label was replaced due to be illegible.No further actions are required at this time.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEYPUMP W/POLE CLMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
15 hampshire street
mansfield, MA 02048
5086183640
MDR Report Key12336486
MDR Text Key267068266
Report Number1282497-2021-10551
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/11/2021
Initial Date FDA Received08/18/2021
Supplement Dates Manufacturer Received08/11/2021
Supplement Dates FDA Received11/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
-
-