TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-900
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation was performed in response to a complaint of error code 4183 on the aia-900 analyzer.The device was being used for diagnosis at the time of the complaint.At the customer site the fse cleaned the x2 sensor because dust was found on the sensor.In conclusion, this indicates that the environmental problem was due to improper routine or preventive maintenance.
 
Event Description
The reported intermittent error 4183, s.Loader -x2 home overrun has been determined to be a reportable malfunction to the fda based on delay in reporting of patient results for class a analytes, per (b)(4) delayed result reportability determination.A field service engineer (fse) was dispatched to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12340126
MDR Text Key267302276
Report Number3004529019-2021-00009
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-