Model Number AIA-900
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation was performed in response to a complaint of error 2076 on the aia-900 analyzer.At the customer site, field service ran a patient sample and was able to reproduce the reported error.During troubleshooting, the fse found the sampler to the dispense lane to be too low and made adjustment.This indicates a failure of the sample nozzle due to the d lane cup requiring an alignment.In conclusion, this investigation confirmed a failure of the sample nozzle to meet specifications or intended use.
 
Event Description
The reported error 2076 on all patient samples has been determined to be a reportable malfunction to the fda based on delay in reporting of patient results for estradiol (e2) and progesterone iii analytes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12340170
MDR Text Key268468384
Report Number3004529019-2021-00010
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-