|
Model Number AIA-360 IMMUNOASSAY ANALYZER REFURBISHED |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by running wash prime and noticing that tube was off from the top of the wash probe.Fse reattaching tube to top of wash probe which resolved the complaint.As a precaution, fse replaced tubes that were bending.Fse validated analyzer by running quality control without error and results passed and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: leak sensor s701 detected is generated when leakage sensor s701 activated.Operator is instructed to contact the service department.The most probable cause of the reported event is due to wash probe feed line popped off top of wash probe.
|
|
Event Description
|
A customer reported getting error 3021 leak sensor s701 detected on the aia-360 analyzer.The customer observed fluid coming from the waste tube and on countertop.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic gonadotropin (bhcg), luteinizing hormone (lhii), estradiol (e2) and follicle stimulating hormone (fsh).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Search Alerts/Recalls
|
|
|