H3, h6: the reported device was received for evaluation.A visual inspection of the returned instrument shows no manufacturing abnormalities.The electrode wires are partially detached.The wand cable and connector are free from any physical damage.Product was out of the original packaging.No packaging returned.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the customer provided image shows a used wand with two wires partially detached on electrode.A clinical review states per the complaint details, it is unknown when the electrode crack occurred in relation to the surgery.Base on the limited information provided, a thorough medical investigation cannot be rendered, nor can a root cause of the reported crack be determined.According to the report, the procedure was completed without significant delay using a back-up device.Since no patient injury or other complications were reported, no further clinical/medical assessment is warranted at this time.Should any other medical information be provided, this complaint would be re-assessed.It was determined the device contributed to the reported event.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|