Catalog Number CLXECP
Device Problems Break (1069); Leak/Splash (1354); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot j148 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot j148 shows no trends.Trends were reviewed for complaint categories, alarm #18: system pressure and centrifuge bowl leak/break.No trends were detected for these complaint categories.Photographs were provided by the customer for evaluation.The complaint kit and smart card were not returned.The provided photographs could not verify the reported centrifuge bowl leak.The centrifuge chamber is not visible in the provided photographs.Review of the photographs show the centrifuge bowl uninstalled from the bowl holder.One of the tabs on the centrifuge bowl is damaged, and there is a scuff on the outside of the centrifuge bowl.The damage to the outer bowl is an indication that the centrifuge bowl dislodged from the bowl holder and impacted something in the centrifuge chamber.A known cause for the centrifuge bowl becoming dislodged from the bowl holder is when the centrifuge bowl is not properly secured into the bowl holder.A material trace of the bowl assembly and its components used to build lot j148 did not find any non-conformances.A device history record review did not identify any related non-conformances, deviations or equipment maintenance events.This kit lot had passed all lot release testing.The centrifuge bowl leak could not be verified based on the photographs provided; therefore, a root cause could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).
 
Event Description
The customer contacted mallinckrodt to report they experienced a centrifuge bowl leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported approximately 10 to 12 minutes into the procedure they received an alarm #18: system pressure alarm and heard a noise.The customer stated they observed a fine spray coming from the centrifuge bowl.The ecp treatment was aborted and no residual blood within the kit was returned to the patient.The customer reported the patient was in stable condition.The customer discarded the kit; however, returned photographs for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
shelbourne building
53 frontage rd suite 300
hampton, NJ 08827
MDR Report Key12341875
MDR Text Key268696901
Report Number2523595-2021-00073
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeAU
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2022
Device Catalogue NumberCLXECP
Device Lot NumberJ148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2021
Initial Date FDA Received08/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight81
-
-