MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 924256, serial/lot #: 082216121a, ubd: 10-jun-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the burr hole cover support clip was seated in the base ring, but the jaws did not close.The surgeon tried reseating the clip multiple times, and also attempted drilling to clear the edge of the burr hole, but this did not work.The support clip was replaced (same base ring was used) and the issue was resolved.No symptoms were reported as a result of the event.
|
|