MEDTRONIC PUERTO RICO OPERATIONS CO. AMPLIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMB1D4 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The actual longevity was less than 80 of the 99.9 longevity limit.The device was fully functional, with no high current drain or evidence of battery problems.Without the history of the programmed settings throughout its service life, there is no way to determine why the longevity did not match the predicted model.Concomitant medical products: 6935m62, lead implanted: (b)(6) 2017; 6935m62 lead implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The cardiac resynchronization therapy defibrillator (crt-d) was explanted and replaced due to the battery reaching elective replacement indicator (eri).The device was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|
|