• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2021
Event Type  malfunction  
Event Description
It was reported that prior to implant, the left ventricular (lv) lead separated following removal of the guidewire.A different lv lead was successfully implanted to resolve the event.The patient was stable and there were no adverse consequences.
 
Manufacturer Narrative
The reported event detachment of device component was confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found the ptfe coating of the stylet was bunched up and clogged inside the inner coil distal to the connector pin and the connector pin and connector cap were pulled out of the connector assembly which is consistent with procedural damage.The cause of the reported event was isolated to the stuck stylet due to bunched up ptfe coating and the connector pin and connector cap pulled out due to excessive force and procedural damage.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
MDR Report Key12348991
MDR Text Key267501221
Report Number2017865-2021-28655
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberS000078152
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-