Model Number 1458Q/86 |
Device Problems
Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to implant, the left ventricular (lv) lead separated following removal of the guidewire.A different lv lead was successfully implanted to resolve the event.The patient was stable and there were no adverse consequences.
|
|
Manufacturer Narrative
|
The reported event detachment of device component was confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found the ptfe coating of the stylet was bunched up and clogged inside the inner coil distal to the connector pin and the connector pin and connector cap were pulled out of the connector assembly which is consistent with procedural damage.The cause of the reported event was isolated to the stuck stylet due to bunched up ptfe coating and the connector pin and connector cap pulled out due to excessive force and procedural damage.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
|
|