MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿ Back to Search Results
Model Number A3059
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: unit received with the mayfield 2 lock having up and down movement and the teeth were grinding when rotated.The disk ratchet and retainer were replaced along with general maintenance and cleaning performed.Root cause: the reported complaint was confirmed.The mayfield 2 lock had up and down movement and the teeth were grinding when rotated, requiring replacement of worn components.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
This is 2 of 2 reports linked to mfg report number 3004608878-2021-00536: a facility reported that the mayfield composite series skull clamp (a3059) was not locking properly when attached to the patient's head.No patient injury or surgical delay has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12356653
MDR Text Key267838477
Report Number3004608878-2021-00537
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Date Manufacturer Received08/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-