Device evaluation visual inspection of the device shows damage at the proximal end of the tip assembly.Visual inspection under a microscope shows damage to the metal coils at the proximal end of the housing, just distal to the cutter window.The cutter returned inside the cutter window and the cutter driver was powered on and the cutter rotates but the cutter will not advance past the damage on the housing.A 0.014¿ guidewire from the lab was front loaded via the distal tip and exited out the proximal end of the gw lumen with no difficulty.Image review one photograph was received from the account for evaluation.The photo shows the distal tip assembly of the hawkone in vitro post usage.Biologics are visible on the tip assembly and damage is noted to the housing at the proximal end just distal to the anchor pockets.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|