ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-04
Device Problems Electrical /Electronic Property Problem (1198); Electro-Static Discharge (2149); Output Problem (3005)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 09/27/2021
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing facial nerve simulation with device use.Programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's activation went well.The recipient's facial nerve stimulation is resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode lead was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed. the device passed some of the electrical tests performed.The device passed the mechanical tests performed.Electrical troubleshooting performed revealed a short from a power node to the case ground at the analog chip.The failure of this device is attributed to a short from a power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key12362689
MDR Text Key268017517
Report Number3006556115-2021-01202
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862250
UDI-Public(01)07630016862250(11)190115(17)220131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2019
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/24/2021
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
Supplement Dates FDA Received09/22/2021
10/21/2021
10/29/2021
02/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age7 YR
-
-