This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No manufacturing or service issues were identified as causes of the customer's reported problem during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Cracked tank cover, broken front cover, and damaged power switch.Started with a visual inspection then filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The reported issue was not duplicated.The power switch was damaged but is still operational.The root cause of the reported issue was found to be design and customer damage from regular use.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.A corrective and preventative action has been opened to address the reported issue.
|