Model Number A2114
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield infinity skull clamp (a2114) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the unit was received with the left and right pawls broken and needing to be replaced.To resolve all issues, the set screw was replaced due to wear along with all broken parts.Additionally, general maintenance and cleaning were performed.Root cause: the reported complaint was confirmed.The ratchet extension was unable to be released due to the left and right pawls being broken.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that the locking mechanism on the mayfield infinity skull clamp (a2114) was not satisfactory.It is unknown if there was patient involvement; however, no patient injury was reported or surgical delay has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Type of Device
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12375604
MDR Text Key268441474
Report Number3004608878-2021-00544
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253761
UDI-Public10381780253761
Combination Product (y/n)N
PMA/PMN Number
K130389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA2114
Device Catalogue NumberA2114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Date Manufacturer Received08/05/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-