The mayfield infinity skull clamp (a2114) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the unit was received with the left and right pawls broken and needing to be replaced.To resolve all issues, the set screw was replaced due to wear along with all broken parts.Additionally, general maintenance and cleaning were performed.Root cause: the reported complaint was confirmed.The ratchet extension was unable to be released due to the left and right pawls being broken.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|