Catalog Number A1059 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
This is 2 of 2 reports linked to mfg report number 3004608878-2021-00547.A facility reported that the locking mechanism on mayfield modified skull clamp (a1059) was very tight and needs to be repaired.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
|
|
Manufacturer Narrative
|
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.The reported complaint was confirmed from the evaluation.Evaluation showed that the lock has rotational and lateral movement and a residue buildup is present.The unit needed repair, replacement of worn parts, general maintenance and cleaning.This device exceeds its expected life of 7 years (manufactured in 2010).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|