Catalog Number 80237
Device Problems Detachment of Device or Device Component (2907); Migration (4003)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/02/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 08/2023).Device not returned.
 
Event Description
It was reported that post an atherectomy procedure of the sfa through the left groin and upon device removal, the tip of the catheter was noted to be missing.Upon x-ray examination, the tip of the catheter and a significant length of the device was still found to be in the artery.The device was attempted to be snared out but the device was stuck.Hence, the patient had to undergo surgery for removal of the device.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.The investigation is inconclusive for the reported detachment and migration issues.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 08/2023), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that post an atherectomy procedure of the sfa through the left groin and upon device removal, the tip of the catheter was noted to be missing.Upon x-ray examination, the tip of the catheter and a significant length of the device was still found to be in the artery.The device was attempted to be snared out but the device was stuck.Hence, the patient had to undergo surgery for removal of the device.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
MDR Report Key12375932
MDR Text Key268431884
Report Number3008439199-2021-00140
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
PMA/PMN Number
K172315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80237
Device Lot Number201155
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/02/2021
Initial Date FDA Received08/27/2021
Supplement Dates Manufacturer Received10/26/2021
Supplement Dates FDA Received10/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-