MAUDE Adverse Event Report: ST PAUL LEVEL ONE; WARMING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL LEVEL ONE; WARMING Back to Search Results
Catalog Number CON-HL-90
Device Problems Overheating of Device (1437); Failure to Calibrate (2440); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Information received a smiths medical fluid warming|level 1 hotline low flow systems - hl-90 goes into over temp and can't calibrate.Additional information received 4 august 2021 (email): issue discovered during testing.No patient involvement.
 
Manufacturer Narrative
H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device presenting wear and tear damaged enclosure, front cover, and line cord, outdated pcb and power switch.The customer stated problem was duplicated.The over temp alarm sounded and stayed on.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE
Type of Device
WARMING
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12380854
MDR Text Key268643166
Report Number3012307300-2021-08936
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-