• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 UNKNOWN ARCTICGEL PADS; UNKNOWN ARCTIC GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 UNKNOWN ARCTICGEL PADS; UNKNOWN ARCTIC GEL PADS Back to Search Results
Catalog Number UNKNOWN
Device Problems Inaccurate Flow Rate (1249); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the arctic sun device did not take up water after emptying the reservoir (b)(4) and the arctic gel pads could not be emptied completely(b)(4).
 
Event Description
It was reported that the arctic sun device did not take up water after emptying the reservoir and the arctic gel pads could not be emptied completely.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be, "flow path is slightly blocked".It was unknown whether the device had met specifications.Based on patient code 2645 the product does not appear to have been used.However, the product is intended to be used for treatment purpose but it was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.Although the product family is unknown, the arctic gel pads ifus are found to be adequate based on past reviews.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ARCTICGEL PADS
Type of Device
UNKNOWN ARCTIC GEL PADS
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12381198
MDR Text Key270166430
Report Number1018233-2021-05229
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/30/2021
Initial Date FDA Received08/28/2021
Supplement Dates Manufacturer Received02/11/2022
Supplement Dates FDA Received03/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-