• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RADIOMETER MEDICAL APS ABL800 FLEX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RADIOMETER MEDICAL APS ABL800 FLEX Back to Search Results
Model Number 393-800
Device Problem High Readings (2459)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2021
Event Type  malfunction  
Event Description
According to the complaint, the customer reported about deviation on k during blood sample measurement on an abl800 on two different patients.No patient was harmed, injured, or maltreated hence no adverse event.Calibration as well as qc showed no problem at all.Patient 1 (female, (b)(6)): sample (b)(4); (b)(6) 2021; 05:17 a.M.; k=11,3 mmol/l value does not match with patient's status.Comparison measurement at customer's lab showed a plausible value k= 3,9 mmol/l.Patient 2 (female, (b)(6)): sample (b)(4); (b)(6) 2021; 07:25 a.M.; k=7,4 mmol/l value does not match with patient's status.Comparison measurement at customer's lab showed a plausible value k= 4,25 mmol/l.
 
Manufacturer Narrative
According to the complaint the issue was resolved by cleaning of measuring chamber and replacement of reference membrane as well as k+ membrane.Based on that the incident is assessed to be caused by a component failure.Due to lack of information it has not been possible to determine the component or root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL800 FLEX
Type of Device
ABL800 FLEX
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA  2700
MDR Report Key12382815
MDR Text Key268739288
Report Number3002807968-2021-00033
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
PMA/PMN Number
K041874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number393-800
Device Catalogue Number393-800
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/20/2021
Initial Date FDA Received08/30/2021
Supplement Dates Manufacturer Received08/31/2021
Supplement Dates FDA Received09/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-