Model Number 8065998162
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Insufficient Information (4580)
Event Date 08/03/2021
Event Type  malfunction  
Event Description
Femto assisted cataract surgery.When the surgeon stepped on the laser pedal to apply the treatment, there was no laser emission.The machine indicated that it was applying the treatment and it produced a treatment record but there were no visible signs of the laser treatment on the patient's eye.The alcon lensx representative was present in the room.He recommended that we re-start the machine.We waited for the machine to do a quick warm up and opened a new patient interface (pi) and contact lens.After restarting the machine and applying the new pi, the laser treatment was successful.However, this caused about 15-minute delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSX
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
ALCON LABORATORIES, INC.
6201 south freeway
fort worth TX 76134
MDR Report Key12391193
MDR Text Key268906978
Report Number12391193
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 08/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8065998162
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/29/2021
Event Location Ambulatory Surgical Facility
Date Report to Manufacturer08/31/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age17520 DA
-
-