Model Number 71440194
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2021
Event Type  malfunction  
Event Description
It was reported that the tip of a genesis ii articular inserter/extractor have a crack.As this was noticed upon inspection, there was not patient involvement.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirms the tip is cracked.This device exhibits signs of significant use and wear.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII ARTICULAR INSERTER/EXTRACT
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12408459
MDR Text Key269405808
Report Number1020279-2021-06674
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010214393
UDI-Public03596010214393
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440194
Device Catalogue Number71440194
Device Lot Number13HM05594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Initial Date Manufacturer Received 08/16/2021
Initial Date FDA Received09/01/2021
Supplement Dates Manufacturer Received09/13/2021
Supplement Dates FDA Received09/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-