The reported event of ¿the device came apart where the two pieces are connected¿ is confirmed.Visual examination of the returned used device found the machined tip detached from the bullseye shaft.Detached tip was not returned for evaluation.No sign of misuse found with returned device.A device history review found no abnormalities that would contribute to this issue.Although this is a reusable device, it is not serviceable; therefore, a service history is not available for review.(b)(4).Per the instructions for use, the user is advised the following: do not use excessive force on instrument to avoid damage or breakage during use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
The sales representative reported on behalf of the customer that the device, kbr178, was being used on (b)(6) 2021 during an unknown procedure and the ¿the unit just comes apart where the two pieces are connected.¿ there was no injury/impact to the patient.After further assessment, it was found that the fragments did fall into the surgical site and were retrieved with a grasper.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|