• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM SC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM SC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45006
Device Problems Mechanical Problem (1384); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2021
Event Type  malfunction  
Event Description
It was reported a loss of rotation occurred.A 1.6mm jetstream sc atherectomy catheter was used over a thruway guidewire in an urgent rotational atherectomy procedure.The jetstream catheter was advanced; however, when reaching a critical point of the lesion, the catheter did not advance, showing a loss of rotation, and consequently was unable to progress completely into the lesion.The physician tried multiple times without success, the results were not as expected.
 
Manufacturer Narrative
Device/media analysis: the jetstream sc-1.6 device was received by boston scientific for analysis.The shaft and the remainder of the device were inspected for damage.Visual examination showed multiple bends on the catheter shaft as well as buckling/kinks located 8.5cm from the tip to proximal 16.5cm.The functionality of the device was checked by setting up the product per the instructions for use (ifu).The device primed as designed.There were no issues with running the device.The device activated and was run for a period of 2 minutes with no issues or errors.Inspection of the remainder of the device revealed no other damage or irregularities.The event of blades not spinning (loss of rotation) was not confirmed.
 
Event Description
It was reported a loss of rotation occurred.A 1.6mm jetstream sc atherectomy catheter was used over a thruway guidewire in an urgent rotational atherectomy procedure.The jetstream catheter was advanced; however, when reaching a critical point of the lesion, the catheter did not advance, showing a loss of rotation, and consequently was unable to progress completely into the lesion.The physician tried multiple times without success, the results were not as expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM SC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road cork
cork IRELA ND
EI   IRELAND
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12413080
MDR Text Key269518828
Report Number2134265-2021-11059
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2022
Device Model Number45006
Device Catalogue Number45006
Device Lot Number0026262249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/13/2021
Initial Date FDA Received09/02/2021
Supplement Dates Manufacturer Received01/19/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-